Techniques for Bulk Material Resistivity Measurements




Written By Mary Anne Tupta Electrical resistivity is a basic property that defines how well a material conducts current. It’s determined by measuring the resistance of a material sample, and then applying geometry factors. The three basic types of bulk materials—metals, insulators, and semiconductors—have different ranges of resistivity. Metals are good conductors of current with typical resistivities of about 10-6 ohm-cm. Insulators are poor conductors with typical resistivities from about 109 to 1020 ohm-cm. Semiconductors conduct current better than insulators but not as well as metals; they may fall anywhere from about 10-3 to 107 ohm-cm. Measuring the Resistivity of Good Conductors. Characterizing a metal’s resistivity requires the measurement of very low resistances (and therefore, low voltages) accurately. The techniques described here can also be used in applications involving other small voltage measurements, such as those needed in measuring the resistance of superconductors, nanowires, carbon nanotubes, graphene (a one-atom-thick form of carbon), and other nanomaterials in which applied power must be kept low to prevent material heating.



Imagine a conductive material sample in the shape of a thin block with thickness t, width w, and some arbitrary length. To find its volume resistivity, a current source is connected at both ends of the sample along its length. Voltmeter leads are placed a known distance apart, L, along the length of its surface. The resistivity can be found by sourcing a known current, I, measuring the voltage drop, V, then calculating the volume resistivity, ρ, from the measured voltage, the magnitude of the source current, the cross-sectional area, A, (=wt), and the distance between the voltmeter leads, L, using the equation: ρ = (V / I) x (wt / L) For metals and other good conductors, the voltage drop is usually just microvolts or nanovolts, so precise measurements are crucial. Potential error sources include test lead resistance, thermoelectric voltages, low frequency noise, external noise sources, Johnson noise, and the use of a voltmeter with insufficient sensitivity. Fortunately, special techniques can reduce the impact of these errors. For example, using the four-wire (Kelvin) measurement method, in which one set of leads are used to source the current and another set are used to measure the voltage drop across the sample, will eliminate the effects of lead resistance.



Thermoelectric voltages are a common source of error when making low voltage and low resistance measurements. These voltages are generated when different metals in the circuit are at different temperatures. To reduce thermoelectric voltages construct test circuits using the same materials for interconnects. Minimize temperature gradients within the test circuit and allow the test equipment to warm up and reach thermal equilibrium. Finally, use an offset compensation method to overcome these unwanted offsets, such as a current-reversal method or the delta mode offset compensation technique. The delta mode technique for removing offsets and low frequency noise involves applying a current and measuring the voltage, then reversing the current and re-measuring the voltage. The difference between the two measurements divided by two is the voltage response of the sample to the applied current level. Repeating the process and using averaging reduces the noise bandwidth and therefore the noise. Although this was once a manual technique, which limited reversal speed to less than 1Hz, newer instruments, such as Keithley’s Model 2182A Nanovoltmeter and Model 6221 Current Source, now automate the technique. This increases the reversal speed, which sets the frequency that dominates the noise. Higher reversal speeds do a better job of removing low frequency noise and thermal drift, because these noise sources have lower power at higher frequencies. External noise sources are interferences created by motors, computer screens, or other electrical equipment. They can be controlled by typical shielding and filtering techniques, or by simply eliminating the noise source. Because these noise sources are often at the power line frequency, avoid test frequencies that are exact multiples or fractions of 60Hz or 50Hz. When using DC instruments and reversal methods, the same result can be achieved by adjusting the instrument’s signal integration period for each measurement to an integer number of power line cycles. In any electrical resistance, thermal energy produces the motion of charged particles. This charge movement results in Johnson noise. Johnson noise may be reduced by reducing the bandwidth using analog or digital filtering, or by reducing the temperature of the device. When considering instruments for resistivity measurements keep in mind that most digital multimeters (DMMs)can’t measure microvolt- or nanovolt-level voltage drops accurately. Generally, an instrument with around 1nV sensitivity is a better choice, which can be found in a good nanovoltmeter. Measuring the Resistivity of Insulators. The techniques used to measure the resistivity of insulators such as paper, rubber, and plastics are very different from those used for conductors. The resistivity of an insulator is determined by applying a voltage to the sample for a specified period of time, measuring the resulting current with an electrometer or picoammeter, then calculating the resistivity based on Ohm’s law and geometric considerations. Both volume and surface resistivity measurements can be made on insulators. Volume resistivity is a measure of the leakage current directly through the insulator. To set up this measurement, picture a sample of arbitrary shape, but with a known uniform thickness t. Two electrodes, each having the same area, A, are placed on the top and bottom of the sample. The lower electrode is connected to the high terminal of a voltage source, V. The low terminal of the voltage source is connected to the low terminal of an ammeter. The high terminal of the ammeter is connected to the top electrode. Although the magnitude of the applied voltage usually depends on the material under test, it is often 500V DC (per ASTM D257). After a specified electrification time, usually 60 seconds, the current, I, is measured using an ammeter capable of measuring nanoamps or lower. Volume resistivity, ρ, is calculated based on the area, A, of the electrodes and the thickness, t, of the sample using the equation: ρ = (V / I) x (A / t) Surface resistivity is the electrical resistance on the sample’s surface. This measurements is conducted on a sample of any thickness and length, but fabricated with a constant width, w. Two electrodes are placed across the width of the sample, separated by a constant distance, L. An ammeter’s high terminal is placed on one electrode; the voltage source’s high terminal is placed on the other. The low terminals of the ammeter and voltage source are connected together. A potential difference, V, is applied for a known period of time and the ammeter measures the resulting current, I. The surface resistivity, o, is calculated from: o = (V / I) x (w / L) Potential error sources when characterizing an insulator’s resistivity include choosing an ammeter without sufficient sensitivity, and using an inappropriate electrification time or test voltage. Other error sources include electrostatic interference, background currents due to charge stored in the material, and static or triboelectric charge, or piezoelectric effects. Electrostatic interference occurs when an electrically charged object is brought near an uncharged object. High resistance materials do not allow the charge to decay quickly and unstable measurements may result. Electrostatic shielding will help to minimize these effects. Shield the material by using a conductive shielded enclosure and connecting the low terminal of the ammeter to the shield. Background currents can equal or exceed the current stimulated by the applied voltage. To counter the effects of these spurious currents, an alternating polarity technique can be used. In this technique, a bias voltage of positive polarity is applied, then the current is measured after a specified delay. Next, the polarity is reversed and the current is re-measured using the same delay. This process can be repeated any number of times. The resistance is calculated based on a weighted average of the most recent current measurements. By calculating a weighted average current, the background current is cancelled out. Some instruments, such as Keithley’s Model 6517B Electrometer, have a built-in test sequence that automates this alternating polarity technique. A high-quality electrode system that provides good contact to the test sample is essential. Conductive rubber on these electrodes enables good contacts to the sample, especially if the sample is a rigid material. Avoid electrodes that will add appreciable resistance to the measurement circuit or could contaminate the sample. Choose an electrode configuration that supports calculating the resistivity from the measured resistance. Several commercial systems now available provide this type of resistivity measurement. Measuring Semiconductor Resistivity The four-point collinear probe technique is the most common way of measuring the resistivity of semiconductor materials, particularly wafers being tested at a probe station. This technique involves the use of four equally spaced (collinear) probes in contact with the material sample. The outer probes (1 and 4) source current; the inner probes (2 and 3) measure the resulting voltage drop across the sample’s surface. The volume resistivity is calculated thus: ρ = (π / ln2) x (V / I) x (tk), where ρ = volume resistivity (ohm-cm), V = voltage between 2 and 3, I = source current from 1 to 4, t = sample thickness (cm), and k = a correction factor based on the ratio of the probe to wafer diameter, and on the ratio of wafer thickness to probe separation. Using four probes in this manner eliminates measurement errors due to the probe and lead resistance, the spreading resistance under each probe, and contact resistance between each metal probe and the semiconductor material. Another technique for measuring the resistivity of semiconductors is the van der Pauw method, which involves applying a current and measuring voltage using four small contacts on the circumference of a flat, arbitrarily shaped sample of uniform thickness. The current is forced between two adjacent terminals of the sample. The voltage is measured on the opposite pair of terminals. This method is particularly useful for measuring very small samples because the dimensions of the sample and the spacing of the contacts are unimportant. It requires making eight measurements around the periphery of the sample to compensate for offsets and geometric considerations. These measurements are combined mathematically to compute the resistivity. Typical sources of error for van der Pauw measurements include voltage drops due to lead and contact resistances, voltage offsets, and incorrect instrumentation choices. In semiconductor material research, a parametric tester is often the instrumentation of choice when characterizing material resistivity. These systems often include a switch matrix to switch the current source and voltmeter to all sides of the sample, which facilitates measurement automation. Parametric test systems also include software to completely automate the measurements and perform resistivity calculations. Special considerations must be taken into account when measuring semiconductor materials with resistances of hundreds of kilo-ohms or higher: * To avoid leakage currents, use a 4-point collinear probe with excellent isolation between the probes to avoid leakage current errors. * Choose a current source with high output impedance (≥1E14 ohms) to avoid loading errors, and with a built-in guard circuit to reduce the effects of shunt capacitance. * Use voltmeters with high input impedance (≥1E14 ohms). * Always shield the sample and all sensitive circuitry, and use shielded cabling to prevent errors due to electrostatic interference; connect the shield to the instrumentation’s low terminal. * To avoid errors from common mode currents, use differential voltage measurement techniques. Summary In such a brief article it is impossible to address all of complexities that help ensure the accuracy of bulk materials resistivity measurements. For more details, view Keithley’s free webinar on this topic at the following link: How to Make Electrical Resistivity Measurements of Bulk Materials. About the Author Mary Anne Tupta is a senior applications engineer at Keithley Instruments, Inc., headquartered in Cleveland, Ohio. She earned a Bachelor of Science degree in physics/electronic engineering and a Master of Science degree in physics from John Carroll University in Cleveland. She has assisted Keithley customers with instrument applications since 1988.


Improving Low Electrical Current Measurements




Characterizing devices at low current levels requires knowledge, skill, and the right test equipment. Even with all three, achieving accuracy in these measurements can be a challenge because the current level is often at or below the noise level of the test setup. To ensure measurement accuracy, it is important to know the type of test equipment to use, the different sources of measurement error, and the appropriate techniques to minimize these errors. Examining several test examples, such as characterization of a field-effect transistor (FET) and a carbon nanotube, can help in the learning process. The term low current is relative, of course. A current level considered low for one application, such as 1mA, may be high for a device operating at 10nA. In general, an instrument’s noise level will establish its low-level sensitivity, with low current measurements referring to those made near an instrument’s noise level. Trends in portable and remote electronic devices, along with advances in semiconductors and nanotechnology, are requiring greater use of low current measurements. Small geometry devices, photovoltaic devices, and carbon nanotubes (CNTs) are a few examples of devices designed to operate at extremely low current levels, and all of these devices must be characterized in terms of their current-voltage characteristics (I-V measurements).



A number of instruments are available for low-current measurements, depending on the type of device under test (DUT) and the level of current to be measured. Perhaps the most ubiquitous tool on production lines and in field service is the digital multimeter (DMM), which typically provides capabilities for measuring current, voltage, resistance, and temperature. The range of commercial products is wide, from low-cost units with 3½-digit readout resolution to rack-mount and benchtop high precision laboratory units. The most sensitive DMMs available can measure current levels as low as about 10pA. When greater precision is needed, various forms of ammeters are available to measure current. These can be as simple as older types that measure current flow from the mechanical deflection of a coil in a magnetic field. More modern digital ammeters use an analog-to-digital converter (ADC) to measure the voltage across a shunt resistor and then determine and display the current from that reading. Newer picoammeters typically use a feedback resistor, which allows more accuracy in current measurements at such low levels. They are available in various configurations, including high-speed models and logarithmic units capable of measuring a wide range of currents. While they are extremely versatile, it is useful to understand the performance limitations of feedback ammeters.



Feedback Ammeter Performance. A simple feedback ammeter can be modeled with a small number of parameters. The current source is modeled as a voltage source in series with a parallel RC circuit, i.e., the source resistance (RS) and parallel source capacitance (CS). The feedback ammeter is modeled as a feedback amplifier with a parallel RC feedback circuit across it (i.e., RF and CF), with the two amplifier inputs being the external current source and the internal voltage noise source, VNOISE. The capacitances in the source and measurement circuits are parasitic elements associated with the resistances and circuit wiring. Using this model and ignoring capacitance, the noise gain of the ammeter circuit can be found from: Output Voltage Noise = (Input Voltage Noise) x (1 + RF/RS) As this equation implies, the output of a feedback ammeter circuit is a voltage, which is proportional to the input current. As the source resistance decreases in value, the output noise increases. When RF = RS, the input noise is multiplied by a factor of 2. If the source resistance is too low, it can have a detrimental effect on the noise performance of the measurement system. The optimum source resistance is a function of required measurement range for an ammeter, with a minimum value of 1MegOhm to measure nanoamps of current, compared to a minimum value of 1GigOhm to measure picoamps of current. However, source capacitance can also affect the noise performance of a low current measurement instrument. In general, as the source capacitance increases, the noise gain also increases. This means that the equation above should be modified by substituting the feedback impedance (ZF) for the feedback resistance (RF) and the source impedance (ZS) for the source resistance (RS). Additional current measurement instruments include electrometers and source-measure units (SMUs). An electrometer is essentially a voltmeter with a high input impedance (1TOhm and higher) that can be used to measure low current levels. It can be used as an ammeter to measure low current levels even at low voltages, and can also be used as a voltmeter to make voltage measurements with minimal effect on the circuit being measured. As an ammeter, an electrometer can measure currents as low as the instrument’s input offset current, as low as 1fA in some cases. As a voltmeter, an electrometer can measure the voltage on a capacitor without significantly discharging the device, and can measure the potential of piezoelectric crystals and high-impedance pH electrodes. The SMU is an innovation for making low-current measurements. It combines precision current sources and voltage sources with sensitive detection circuitry for measuring both current and voltage. An SMU can simultaneously provide a source of current and measure voltage or provide a source of voltage and measure current. A well-equipped SMU may include a voltage source, current source, ammeter, voltmeter, and ohmmeter and is also programmable for use in automatic-test-equipment (ATE) systems. Minimizing External Noise. All of these measuring instruments are effective tools for measuring current, but their sensitivity to low levels of current will be limited mainly by sources of noise, both within and external to the instrument. The DUT also affects the level of current that can be accurately measured with a given instrument, because the DUT’s source resistance (RS) establishes the level of Johnson current noise (IJ), which is low-level noise caused by temperature effects on electrons in a conductor. Johnson noise, which can be expressed in terms of either current or voltage, is essentially the voltage noise of a device divided by the device resistance: IJ= √(4kTB/RS) / RS, wherek = Boltzmann’s constant (1.38 × 10–23 J/K), T = Absolute temperature of the source (in ºK), B = the noise bandwidth (in Hz), and RS = the resistance of the source (in ohms). Both temperature and noise bandwidth affect the Johnson current noise. A reduction in either parameter will also reduce the Johnson current noise. Cryogenic cooling, for example, is often used to reduce noise in amplifiers and other circuits but adds cost and complexity. The noise bandwidth can be reduced by filtering, but this will result in slowing the measurement speed. The Johnson current noise also decreases as the DUT’s source resistance decreases, but this is not often a practical or even possible option. Ideally, a current measurement would be just that of the DUT source. However, current noise from various unwanted sources can make it difficult to read a low-level DUT source current. One of these unwanted sources is part of the measurement system itself, i.e., the coaxial cables used to interconnect test instruments to each other or to the DUT. Typical test cables can generate as much as tens of nanoamps of current as a result of the triboelectric effect. This occurs when the outer shield of a coaxial test cable rubs against the cable’s insulation when the cable is flexed. As a result, electrons are stripped from the insulation, and added to the current total. In some applications, such as nanotechnology and semiconductor research, the current generated by this effect may exceed the level of current to be measured from the DUT. Triboelectric effects can be minimized by using low-noise cable, with an inner insulator of polyethylene coated with graphite underneath the outer shield. The graphite reduces friction, and provides a path for the displaced electrons to return to their original locations, eliminating random electron motion and their contribution to the additional noise level. Excess current flow from triboelectric effects can also be minimized by reducing the length of the test cables as much as possible. The test setup should be isolated from vibration to minimize unwanted movement of the test cables, by positioning test cables on top of vibration-absorbing material, such as foam rubber. Test cable movement can also be minimized by taping the cables to a stable surface, such as the test bench. Piezoelectric effect is another source of error in low-current measurements. It causes spurious current generation due to mechanical stress on susceptible materials. The effect varies by material, although some materials commonly used in electronic systems, such as polytetrafluoroethylene (PTFE) dielectrics, can produce a relatively large amount of current for a given amount of stress and vibration. Ceramic materials are less affected by piezoelectric effects and produce lower current levels. To minimize current generated by this effect, it is critical to minimize mechanical stress on insulators and construct the low-current test system using insulating materials with minimal piezoelectric properties. Insulators can also degrade low-current measurement accuracy by means of dielectric absorption. This phenomenon occurs when a high-enough voltage across an insulator causes positive and negative charges to polarize. When the voltage is removed from the insulator, it gives up the separated charges as a decaying current, which is added to the total amount measured during a test. The decay time for the current from dielectric absorption to dissipate can be from minutes to hours. The effect can be minimized by applying only low-voltage levels to insulators used for low-current measurements. Insulators can also degrade low-current measurement accuracy due to contamination from salt, moisture, oil, or even fingerprints on the surface of the insulator. Contamination effects can also plague printed circuit boards in a test fixture or in the test setup when, for example, excessive flux is used when soldering. On an insulator, the contamination acts to form a low-current battery at a sensitive current node within the insulator, generating noise currents that can be on the order of nanoamps. To minimize measurement errors from insulator contamination, an operator should wear gloves when handling insulators or simply avoid touching them. The use of solder should be minimized, and solder areas should be cleaned with an appropriate solvent, such as isopropyl alcohol. A clean cotton swab should be used for every cleaning, and cotton swabs should never be reused or dipped into the cleaning solution after having been used for cleaning. It is critical to make low-current measurements in the absence of magnetic fields, because such fields can induce current flow in conductors. This is typically due to variations in magnetic field intensity, or motion of a conductor within a magnetic field. Both cases should be avoided to maintain measurement accuracy, which is best accomplished by properly shielding the measuring instrument or system. Minimizing Instrument Offset Current. An instrument used for low-current measurements should show a zero reading when its input terminals are left in an open-circuit condition. Unfortunately, this is rarely the case due to a small current known as the input offset current. It is caused by bias currents of active devices in measuring instrument circuitry, as well as leakage current through insulators in the instrument or test system. Most instrument manufacturers specify the input offset current on their products’ data sheets for comparison purposes, and this small amount of current must be taken into account in any low-current measurement. In other words, the instrument’s reading is actually the sum of the DUT source current and the instrument’s input offset current. The input offset current can be found by capping the input connector and selecting the lowest current range available on the measuring instrument. The reading shown by the instrument, after it has properly settled to a stable value, should be within the specification shown on the instrument’s data sheet and can be subtracted from DUT readings. On some instruments, a current-suppression function can partially null input offset current. Another way to subtract input offset current from a low-current measurement is to use a relative function found on some measuring equipment, such as ammeters. The relative function stores the reading of whatever residual offset current is being measured with the input terminals left in an open-circuit condition; this reading is treated as the zero point for subsequent readings. Application Examples. Some examples of practical low-current measurements include characterization of field effect transistors (FETs) and CNT devices. A more common FET test involves evaluation of a device’s common-source characteristics. Even at low current levels, the drain current can be studied using a simple test setup with a two-channel SMU, such as the Keithley Series 2600A System SourceMeter instrument. A two-channel SMU has the capability to source current or voltage and measure current or voltage simultaneously. To characterize a FET, it is mounted in a test fixture that allows secure ground and bias connections. One SMU channel supplies a swept gate-source voltage (VGS) to the FET while the other supplies a swept drain-source voltage (VDS) and measures the FET’s drain current (ID). This simple test setup allows the measurement of drain currents as low as 10nA or less. Electronic materials such as photovoltaic wafers and CNT sheets are typically characterized in terms of their current density—the amount of current they can generate for a given area of material. Researchers from South Korea’s Seoul National University, conduct such tests to evaluate multi-walled carbon nanotube (MWNT) devices fabricated on an arc-discharge CNT substrate using a Keithley Model 6517 electrometer [1]. In these studies, current densities as low as 10–4/cm2 were measured at applied electric fields of 5V/μm and less. Practical analysis of the I-V characteristics of CNT-based electronics can also be performed in a manner similar to that for the FET by using a pair of SMUs to sweep drain and gate voltages while measuring and plotting the drain current as a function of gate voltage. The required resolution and accuracy of low-current measurements will dictate the type of measurement tool used. When accuracy is less of an issue, a basic DMM may suffice. But for more demanding requirements, a precision electrometer or SMU may be needed. These precision instruments are optimized for low-current measurements, providing measurement resolution as small as 1fA. More techniques and tips on low current measurements are contained in Keithley’s Low Level Measurements Handbook [2]. References. 1. Joeoong Hahn, Jae-Eun Yoo, Jaeik Han, Hyok Bo Kwon, and Jung Sang Suh, "Field emission from the film of the finely dispersed arc discharge black core material," Carbon, Vol. 43, 2005, pp. 937-943. 2. Keithley Instruments, Inc., "Low Level Measurements Handbook, 6th Edition", 2004, available for download at http://gw1.vtrenz.net/?SIP0AWDVFA\.


Friday, 13 April 2012

Fundamentals of Semiconductor C-V Measurements




A Universal Test


Capacitance-voltage (C-V) testing is widely used to determine semiconductor parameters, particularly in MOSCAP and MOSFET structures. However, other types of semiconductor devices and technologies can also be characterized with C-V measurements, including bipolar junction transistors (BJTs), JFETs, III-V compound devices, photovoltaic cells, MEMs devices, organic TFT displays, photodiodes, carbon nanotubes (CNTs), and many others.The fundamental nature of these measurements makes them useful in a wide range of applications and disciplines. They are used in the research labs of universities and semiconductor manufacturers to evaluate new materials, processes, devices, and circuits. C-V measurements are extremely important to product and yield enhancement engineers, who are responsible for improving processes and device performance. Reliability engineers use these measurements to qualify material suppliers, monitor process parameters, and analyze failure mechanisms.With appropriate methodologies, instrumentation, and software, a multitude of semiconductor device and material parameters can be derived. This information is used all along the production chain beginning with evaluation of epitaxially grown crystals, including parameters such as average doping concentration, doping profiles, and carrier lifetimes. In wafer processes, C-V measurements can reveal oxide thickness, oxide charges, mobile ions (contamination), and interface trap density. These measurements continue to be used after other process steps, such as lithography, etching, cleaning, dielectric and polysilicon depositions, and metallization. After devices are fully fabricated on the wafer, C-V is used to characterize threshold voltages and other parameters during reliability and basic device testing and to model the performance of these devices.The Physics of Semiconductor Capacitance


A MOSCAP structure is a fundamental device formed during semiconductor fabrication. Although these devices may be used in actual circuits, they are typically integrated into fabrication processes as a test structure. Since they are simple structures and their fabrication is easy to control, they are a convenient way to evaluate the underlying processes.The metal/polysilicon layer is one plate of the capacitor, and silicon dioxide is the insulator. Since the substrate below the insulating layer is a semiconducting material, it is not by itself the other plate of the capacitor. In effect, the majority charge carriers become the other plate. Physically, capacitance, C, is determined from the variables in the following equation:C = A (?/d), where A is the area of the capacitor, ? is the dielectric constant of the insulator, and d is the separation of the two plates.Therefore, the larger A and κ are, and the thinner the insulator is, the higher the capacitance will be. Typically, semiconductor capacitance values range from nanofarads to picofarads, or smaller.The procedure for taking C-V measurements involves the application of DC bias voltages across the capacitor while making the measurements with an AC signal. Commonly, AC frequencies from about 10kHz to 10MHz are used for these measurements. The bias is applied as a DC voltage sweep that drives the MOSCAP structure from its accumulation region into the depletion region, and then into inversionA strong DC bias causes majority carriers in the substrate to accumulate near the insulator interface. Since they can't get through the insulating layer, capacitance is at a maximum in the accumulation region as the charges stack up near that interface (i.e., d is at a minimum). One of the fundamental parameters that can be derived from C-V accumulation measurements is the silicon dioxide thickness, tox.As bias voltage is decreased, majority carriers get pushed away from the oxide interface and the depletion region forms. When the bias voltage is reversed, charge carriers move the greatest distance from the oxide layer, and capacitance is at a minimum (i.e., d is at a maximum). From this inversion region capacitance, the number of majority carriers can be derived. The same basic concepts apply to MOSFET transistors, even though their physical structure and doping is more complex.Many other parameters can be derived from the three regions as the bias voltage is swept through them. Different AC signal frequencies can reveal additional details. Low frequencies reveal what are called quasistatic characteristics, whereas high frequency testing is more indicative of dynamic performance. Both types of C-V testing are often required.Basic Test Setup


Because C-V measurements are actually made at AC frequencies, the capacitance for the device under test (DUT) is calculated with the following:CDUT = IDUT / 2?fVAC, where IDUT is the magnitude of the AC current through the DUT, f is the test frequency, and VAC is the magnitude and phase angle of the measured AC voltageIn other words, the test measures the AC impedance of the DUT by applying an AC voltage and measuring the resulting AC current, AC voltage, and impedance phase angle between them. These measurements take into account series and parallel resistance associated with the capacitance, as well as the dissipation factor (leakage).Challenges to Successful C-V MeasurementsCertain challenges are associated with this testing. Typically, test personnel have problems in the following areas:


Low capacitance measurements (picofarads and smaller values)


C-V instrument connections (through a prober) to the • wafer device


Leaky (high D) capacitance measurements


Using hardware and software to acquire the data


Parameter extractionsOvercoming these challenges requires careful attention to the techniques used along with appropriate hardware and software.Low Capacitance Measurements. If C is small, the DUT's AC response current is small and hard to measure. However, at higher frequencies, the DUT impedance is reduced, so the current increases and is easier to measure. Often semiconductor capacitance is very low (less than 1pF), which is below the capabilities of many LCR meters. Even those claiming to measure these small capacitance values may have confusing specifications that make it difficult to determine the final accuracy in the measurement. If accuracy over the instrument's full measurement range is not explicitly stated, the user needs to clarify this with the manufacturer.High D (Leaky) Capacitors. In addition to having a low C value, a semiconductor capacitor may also be leaky. That is the case when the equivalent R in parallel with C is too low. This results in resistive impedance overwhelming the capacitive impedance, and the C value gets lost in the noise. For devices with ultra-thin oxide layers, D values can be greater than five. In general, as D increases, the accuracy of a C measurement is rapidly degraded, so high D is a limiting factor in the practical use of a C meter. Again, higher frequencies can help solve the problem. At higher frequencies the capacitive impedance is lower, resulting in a C current that is higher and more easily measured.C-V Measurement Connections. In most test environments, the DUT is a test structure on a wafer: It is connected to the C-V instrument through a prober, a probe card adapter, and a switch matrix. Even if no switch is involved, there is still a prober and significant cabling. At high frequencies, special corrections and compensation must be applied. Usually, this is achieved with some combination of an open, short, or calibration device. Because of the complexity of the hardware, cabling, and compensation techniques, it is a good idea to confer with C-V test application engineers. They are skilled at working with various probe systems to overcome many types of interconnection problems.Obtaining Useful Data. In addition to the accuracy issues mentioned earlier, practical considerations in C-V data collection include the instrumentation's range of test variables, versatility of parameter extraction software, and ease of hardware usage. Traditionally, C-V testing has been limited to about 30V and 10mA DC bias. However, many applications, such as characterizing LD MOS structures, low-kinterlayer dielectrics, MEMs devices, organic TFT displays, and photodiodes, require tests at higher voltage or current. For these applications, a separate high voltage DC power supply and C meter are required; DC bias up to 400V differential (0 to +/-400V) and a current output up to 300mA are very useful. Being able to apply differential DC bias on both the HI and LO terminals of the C-V instrument offers more flexible control over electric fields within the DUT, which is very helpful in the research and modeling of novel devices, such as nanoscale components.The instrumentation software should include ready-to-run test routines that do not require user programming. These should be available for the most widely used device technologies and test regimens, which were mentioned in the first three paragraphs of this article. Some researchers may also be interested in less common tests, such as performing both a C V and C f sweep on a Metal Insulator Metal (MIM) capacitor, measuring small interconnect capacitance on a wafer, or doing a C V sweep on a two-terminal nanowire device. The parameter extractions should be easily obtained, with automated curve plotting.Often, engineers and researchers are expected to perform C-V measurements with little experience and training on the instrumentation. A test system with an intuitive user interface and easy-to-use features makes this practical. That includes simple test setup, sequence control, and data analysis. Otherwise, the user spends more time learning the system than collecting and using the data. Other considerations are a test system with:Tightly integrated source-measure units, digital oscilloscope and C-V meter


Easy integration with other external instruments


High resolution and precise measurements at the probe tips (DC biasing down to millivolts and capacitance measurements down to femtofarads)


Test setups and libraries that can be easily modified


Diagnostic/troubleshooting tools that let users know whether or not the system is performing correctly.